Model Number 20188
Device Problems Device Dislodged or Dislocated (2923); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/30/2018
Event Type  malfunction  
Event Description
It was reported that stent dislodgement occurred.The target lesion was located in the heavily calcified left renal artery.A 6.0mmx18mmx150cm express sd renal/biliary stent was advanced to treat the lesion; however, difficulty advancing was encountered.The physician exerted much pressure causing the delivery system shaft to bend at the level of the stent.The stent advanced partially across the ostium of the renal artery.Upon removal of the un-deployed stent delivery system (sds), it was noted that the stent separated from the balloon shaft but remained on the wire.The sds was advanced up to level of stent and partial recapture accomplished.The stent was pulled back into the sheath and the entire sds, sheath, and the stent were removed from the patient's body.The procedure was ended and the patient will return for the same procedure, but using a different approach.No patient complications were reported and the patient was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS SD RENAL/BILIARY
Type of Device
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8092419
MDR Text Key128030661
Report Number2134265-2018-63047
Device Sequence Number1
Product Code NIN
UDI-Device Identifier08714729484844
UDI-Public08714729484844
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model Number20188
Device Catalogue Number20188
Device Lot Number0022454778
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/30/2018
Initial Date FDA Received11/20/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age78 YR
-
-