Model Number PHA0-1252
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Type  Injury  
Manufacturer Narrative
This complaint will be updated once the investigation is complete.Trends will be evaluated.
 
Event Description
Allegedly, modular neck revised due to a fracture in the neck note don imaging.Upon revision, the surgeon noted that the neck fracture was due to a bipolar shell (depuy) that was utilized in a prior procedure in a "tripolar" capacity.The bipolar shell was resting on the neck at the site of the fracture and is thought to have initialized the modular neck fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFEMUR® MODULAR FEMORAL NECK
Type of Device
HIP COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer (Section G)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key8094929
MDR Text Key128050552
Report Number3010536692-2018-01463
Device Sequence Number1
Product Code LWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K003016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberPHA0-1252
Device Catalogue NumberPHA0-1252
Device Lot Number028550129
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date11/06/2018
Date Manufacturer Received11/06/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age64 YR
-
-