A technical analysis of the returned instrument was conducted and are view of the manufacturing history was performed to evaluate whether there was any deviation that could account for the reported event.The instrument under complaint was returned completely disassembled with the inner and outer shaft separated.Several kinks were observed in both shafts.An approximately 1.5 cm long fragment of the stent was returned.The outer shaft, which has to be retracted in order to release the stent, was most likely fractured inside the handle.The inner shaft was found ruptured approximately 3 mm distal to the metal tube where the knife is mounted on.Based on the conducted investigations no manufacturing or material related root cause could be identified.The blockage of the stent is most likely related to uncommonly high friction between the stent and the retractable outer shaft.
|