Investigation results: 10-may-2018 on e801 module/ cobas e 411 immunoassay analyzer (e411) serial number (b)(4), ft3 iii (pg/ml): 8.97/ 5.08, ft4 iii (ng/dl): 1.39/ 1.38, tsh (uiu/ml): 2.40/ 2.3.11-oct-2018 on e801 module/ e411, ft3 iii (pg/ml): 6.38/ 4.01, ft4 iii (ng/dl): 1.29/ 1.26, tsh (uiu/ml): 3.43/ 3.07.The calibration and qc data for the investigation are ok.A general reagent issue can be excluded.The investigation did not identify a product problem.The cause of the event could not be determined.
|