• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM AC POWER SUPPLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM AC POWER SUPPLY Back to Search Results
Catalog Number 295050-001
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because this issue did not prevent the freedom driver from performing its life-sustaining functions.In addition, the freedom driver has a redundant power source of onboard batteries.The freedom ac power supply has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that the freedom ac power supply was not working; the led light did not illuminate when inserted in the patient's freedom driver.There was no reported adverse patient impact.
 
Manufacturer Narrative
The customer-reported issue of the freedom ac power supply not able to supply power to the freedom driver was confirmed during the functional evaluation; the ac power supply was unable to pass current to the driver.The root cause of the reported issue cannot be determined without physical access to the internal printed circuit board.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM AC POWER SUPPLY
Type of Device
AC POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key8113014
MDR Text Key129176365
Report Number3003761017-2018-00542
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 01/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295050-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2018
Date Manufacturer Received11/19/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age21 YR
-
-