Returned for evaluation was an opened duramax catheter.A visual examination of the returned device noted the catheter was fractured into 2 pieces, with the site of the fracture located approximately 1" below the bifurcate suture wing.The catheter exhibited a pinch mark at the fracture site.The customer's reported complaint of "the catheter fractured completely during insertion" is confirmed, however a definitive root cause cannot be determined.A review of the lot history records was performed for the packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use, provided to the end user with this product, states: "do not use sharp instruments near the extension tubing or catheter lumen.Do not use scissors to remove dressing.Catheter will be damaged if clamps other than what is provided with this kit are used.Examine catheter lumen and extensions before and after each treatment for damage.Caution: do not clamp the dual lumen portion of the catheter.Clamp only the extensions.Do not use serrated forceps, use only the in-line clamps provided." a review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
|