• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC0301
Device Problem Leak/Splash (1354)
Patient Problem Bradycardia (1751)
Event Date 11/08/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).The customer reported the device was discarded.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report the leak and treatment with aspiration and medication.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.The steerable guide catheter (sgc) was advanced to the left atrium.The dilator was removed for aspiration and air was observed in the syringe.It was confirmed that the sgc was not against any structures of the heart.Aspiration was performed and the air was removed.The clip delivery system (cds) was advanced to the mitral valve.While grasping, the patients heart rate suddenly dropped to 13 beats per minute (bpm).Atropine was administered and the patient was stabilized.The clip was deployed without issue, reducing mr to 1-2.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Internal file number - (b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no similar incidents reported from this lot.All available information was investigated and a definitive cause for the reported in leak in this incident could not be determined.The definitive cause for bradycardia could not be confirmed.The reported patient effect of bradycardia is listed in the mitraclip system instructions for use as a known possible complication associated with mitraclip procedures.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8117192
MDR Text Key128860909
Report Number2024168-2018-09149
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/12/2018
Device Catalogue NumberSGC0301
Device Lot Number80730U135
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/08/2018
Initial Date FDA Received11/29/2018
Supplement Dates Manufacturer Received02/21/2019
Supplement Dates FDA Received02/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age80 YR
Patient Weight118
-
-