SMITHS MEDICAL ASD, INC. PORTEX® EPIDURAL MINIPACKS; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number EPX995/118BJP
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
Information was received that a smiths medical catheter had wet dressing upon removal, two days after conducting operation.An accidental removal of the catheter occurred.No patient injury resulted.
 
Manufacturer Narrative
Evaluation summary: one epifuse connector was returned for investigation.After visually inspecting the sample, no damage was seen.After unlocking the epifuse connector, a catheter removal test was performed according to the product standard, but catheter removal was not confirmed.A device history record review was also completed and no discrepancies were found.Based on the investigation and record review, the complaint allegation was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® EPIDURAL MINIPACKS
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
avenida calidad #4
parque industrial internationa
tijuana, 22380
MX   22380
Manufacturer Contact
dave halverson
6000 nathan lane north
minneapolis, MN 55442
7633833310
MDR Report Key8122659
MDR Text Key129028461
Report Number3012307300-2018-08596
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberEPX995/118BJP
Device Lot Number180808
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received12/06/2018
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-