MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ELUVIATM DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ELUVIATM DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24620
Device Problem Fracture (1260)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/16/2018
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.
 
Event Description
It was reported that there was a stent fracture.The patient was enrolled in the imperial study on (b)(6) 2016, and the index procedure was performed on the same day.The target lesion was located in the left proximal superficial femoral artery (sfa).The target lesion had a vessel diameter of 6mm, both proximally and distally, was 100mm long, and was 100% stenosed.The target lesion was treated with pre-dilatation, and a 7x120mm eluvia stent was placed.There was 0% residual stenosis.On (b)(6) 2018, during the two year protocol follow up, an x-ray revealed a stent fracture.There was no further action taken.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIATM DRUG-ELUTING VASCULAR STENT SYSTEM
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8125286
MDR Text Key129065857
Report Number2134265-2018-63382
Device Sequence Number1
Product Code NIU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P180011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2017
Device Model Number24620
Device Catalogue Number24620
Device Lot Number18348447
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/21/2018
Initial Date FDA Received12/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
-
-