Catalog Number 180503
Device Problem Osseointegration Problem (3003)
Patient Problems Pain (1994); Injury (2348); Inadequate Osseointegration (2646)
Event Date 11/05/2018
Event Type  Injury  
Manufacturer Narrative
An event regarding loosening involving a mako femoral component was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Medical records received and evaluation: not performed as medical records were not provided.Device history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: an event regarding loosening involving a mako uni femoral component was reported.The exact cause of the event could not be determined because insufficient information was provided.Additional information, including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event Description
It was reported that patient's knee was revised due to loosening after patient complaint of pain.Update: the femoral component was slightly loose.The tibial component and spacer were well-fixed and well-positioned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCK FEMORAL-LM-RL-SZ 3
Type of Device
KNEE JOINT PATELLOFEMOROTIBIAL POLYMER/METAL/POLYMER SEMI-CONSTRAINED CEMENTED
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key8126620
MDR Text Key129191849
Report Number3005985723-2018-00701
Device Sequence Number1
Product Code NPJ
UDI-Device Identifier00848486000554
UDI-Public00848486000554
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/30/2020
Device Catalogue Number180503
Device Lot Number457821-M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/05/2018
Initial Date FDA Received12/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/05/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age54 YR
-
-