• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY PROFESSIONAL MIDWEST STYLUS ATC HANDPIECE; HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY PROFESSIONAL MIDWEST STYLUS ATC HANDPIECE; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 882300
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.Evaluation found cap getting warm while running with no load.Button ok and not sticking, no contact with chuck pusher.Bearings noisy and rpm's low.Both bearings failing.
 
Event Description
It was reported that a stylus atc mini handpiece overheated during use; no injury resulted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST STYLUS ATC HANDPIECE
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
901 west oakton st.
des plaines IL 60018
Manufacturer (Section G)
DENTSPLY PROFESSIONAL
901 west oakton st.
des plaines IL 60018
Manufacturer Contact
karl nittinger
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494424
MDR Report Key8135413
MDR Text Key129468497
Report Number9614977-2018-00191
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K003518
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number882300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/05/2018
Initial Date FDA Received12/05/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-