No product samples were returned for investigation, however, a diagram was provided that described the leak coming from the clave portion of the vented bag set assembly.The device history record for lot 3600944 was reviewed and there were no non-conformances found that would have contributed to the reported complaint.Without the actual complaint sample it is not possible to determine a root cause for the reported leakage.
|
The event involved a vented bag spike, clave that leaked at the clave connection during set-up of chemotherapy before infusion to the patient.The device was replaced with no further problems encountered.There was no patient involvement, no serious injury, no medical intervention required, no blood loss nor chemo exposure.
|