The device was examined by a draeger field service engineer.It could be revealed that the ceramic disc of the inspiratory valve was chipped.Due to the missing fragment and the resulting sharp contour the valve disc was jammed inside the cage.The inspiratory path thus remained open during inspiration; a sufficient ventilation was not possible anymore.From the readings and curves on the display as well as from the alarms the user was able to derive that the applied tidal volumes were significantly below the settings.A clear cause for the breaking of the valve could not be established.It is however seen unlikely that the disc can get damaged during its operational movement.It would be rather associated to inadequate reprocessing or false reassembly after that.Checking of the valves is part of the manual checklist to be performed prior to each use; for that purpose the cover of the breathing system is made from transparent material that the user can check if all valve discs are in place and free of damages.The particular valve disc had been replaced in follow-up of the event; the device passed all tests and is back in use.
|