Event Date 11/13/2018
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an unknown specialty therapy device leaked from an unspecified location.This was identified prior to patients use.The device was filled with fluorouracil.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The lot number was manufactured between july 27, 2018 - july 31, 2018.The actual device was received for evaluation.Visual inspection on the returned sample revealed fluid inside the bag that contained the sample.The cause of the fluid inside the bag was due to the untightened blue winged luer cap.Functional testing was performed by filling the device with colored water.After fill, the blue winged luer cap was hand tightened.The sample was monitored until the next day and no signs of leak were observed.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
irvine CA
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8142614
MDR Text Key129747501
Report Number1416980-2018-07682
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 01/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C4711K
Device Lot Number18G033
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/02/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-