Model Number EP003994S
Device Problems Break (1069); Activation, Positioning or Separation Problem (2906)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a radiofrequency procedure, the needle had scraped the lumen of the catheter.The needle was replaced with a competitor needle with resolve.The case was completed with radiofrequency.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the needle, ep003994s with lot number 212174649, was returned and analyzed.The needle was visually inspected, and no anomaly was noticed.In conclusion, the reported skiving issue was not confirmed.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary: the transseptal needle, (b)(4) with lot number 212174649, was returned and analyzed.Visual inspection of the needle showed that no issue was identified.The reported skiving issue was not confirmed through visual inspection of the product.However, it is plausible that skiving occurred as there is a compatibility issue with the needle and a competitor sheath.The cause of the issue was not established.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROCKENBROUGH NEEDLE
Type of Device
TROCAR
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8148219
MDR Text Key129899679
Report Number9612164-2018-03541
Device Sequence Number1
Product Code DRC
UDI-Device Identifier00613994663146
UDI-Public00613994663146
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/16/2018
Device Model NumberEP003994S
Device Catalogue NumberEP003994S
Device Lot Number212174649
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2018
Initial Date FDA Received12/10/2018
Supplement Dates Manufacturer Received01/21/2019
03/05/2020
Supplement Dates FDA Received02/08/2019
03/16/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-