Various electrical loads were attached to the pulse generator and results of diagnostic tests demonstrate that accurate resistance measurements were obtained in all instances.The pulse generator diagnostics were as expected for the programmed parameters.There were no performance or any other type of adverse conditions found with the pulse generator.An analysis was performed on the returned lead portion and the reported allegations were not confirmed.A large portion of the lead assembly (body) including the electrodes was not returned for analysis and therefore a complete evaluation could not be performed on the entire lead product.What appeared to be white deposits were observed in various locations.Eds (energy dispersion spectroscopy ¿ provides chemical or element identity/composition analysis) was performed on the deposit observed on the bilumen tubing and identified the deposit as containing silicon, phosphorus, sodium, magnesium and calcium.The condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.Continuity checks of the returned lead portion were performed, during the visual analysis, and no discontinuities were identified.Based on the findings in the lab, there is no evidence to suggest discontinuities in the returned portion of the device.
|