Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/21/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Foreign - (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during a hip surgery, the crew couldn't be inserted until the end, and couldn't be fixed.The surgery was completed with another screw.No adverse events have been reported as a result of the malfunction.
|
|
Manufacturer Narrative
|
(b)(4).Udi : (b)(4).This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was not confirmed.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.Visual inspection of the screw identified the scratches near the head.There were scratches around the surface of the screw indicating the screw attempt to impact in the shell.The dimensional analysis was performed and was conforming to print specifications.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|