• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES(IRELAND) CUSA EXCEL 36KHZ STRAIGHT HANDPIECE; ULTRASONIC SURGICAL PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES(IRELAND) CUSA EXCEL 36KHZ STRAIGHT HANDPIECE; ULTRASONIC SURGICAL PRODUCTS Back to Search Results
Catalog Number C2602
Device Problem Failure to Power Up (1476)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/14/2018
Event Type  malfunction  
Manufacturer Narrative
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
 
Event Description
A customer reported that a c2602 cusa excel 36khz straight handpiece was not working during a procedure.They tried using a different handpiece and it worked.Additional information received on (b)(6) 2018 indicated that the event happened on (b)(6) 2018 during a craniotomy (brain tumor) case.The device did not cavitate.The device was in contact with the patient but no injury was reported.There was a 15-minute surgery delay but there was no impact to the patient.The case was completed with another handpiece.Additional information has been requested.
 
Event Description
N/a.
 
Manufacturer Narrative
The device was returned for evaluation.The device history record (dhr) reveals no anomalies that could be associated with the complaint incident were observed.A defective coilform was verified as the cause of the complaint incident.Further additional information regarding the failure and cause was not provided.The complaint was verified as valid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSA EXCEL 36KHZ STRAIGHT HANDPIECE
Type of Device
ULTRASONIC SURGICAL PRODUCTS
Manufacturer (Section D)
INTEGRA LIFESCIENCES(IRELAND)
ida business&technology park
ida business&technology park
sragh, tullamore, co.offaly
EI 
MDR Report Key8152750
MDR Text Key130348073
Report Number3006697299-2018-00077
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
PMA/PMN Number
K141674
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC2602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2018
Date Manufacturer Received12/17/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-