Model Number 595000-001
Device Problem Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the reported issue did not prevent the freedom driver from performing its life sustaining function.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia authorized distributor, reported that the freedom driver made a strange noise while supporting a patient.The customer also reported that the patient was switched to a backup driver without any adverse patient impact.
 
Manufacturer Narrative
The customer-reported issue, a strange sound coming from the driver, was reproduced during the extended observation run of the driver.The root cause was determined to be excessive wear of the piston cylinder assembly's (pca's) yoke dowel pin on the pca's cam follower.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.Ce 4591 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key8154939
MDR Text Key130049157
Report Number3003761017-2018-00554
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 02/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2018
Initial Date Manufacturer Received 11/27/2018
Initial Date FDA Received12/12/2018
Supplement Dates Manufacturer Received11/27/2018
Supplement Dates FDA Received02/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age42 YR
-
-