There was no patient involvement.Livanova (b)(4) manufactures the s3 console.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was not able to confirm the reported issue.The device was intensively tested but the reported issue could not be confirmed or reproduced.However as the reported issue could be intermittent the technician decided to replace the dc/dc module.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|