Model Number D224TRK
Device Problems Signal Artifact/Noise (1036); Electromagnetic Interference (1194); Over-Sensing (1438); Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Non specific EKG/ECG Changes (1817); Dizziness (2194); Shock from Patient Lead(s) (3162)
Event Date 08/03/2014
Event Type  Injury  
Manufacturer Narrative
This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that patient felt an electrical shock.Device interrogation noted external electromagnetic interference on both leads and an inappropriate shock was delivered.It was suspected that the patient touched a power source that was not properly grounded.The patient reported feeling ¿slightly dizzy¿ immediately before the shock.This was reportedly due to the patient¿s pacemaker dependency and the inhibition of pacing by the device lead to a brief period of asystole.The device was reprogrammed and remains in use.The patient is a participant in the post approval clinical surveillance product surveillance registry.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSULTA CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key8155693
MDR Text Key130071310
Report Number3004209178-2018-27460
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00613994446626
UDI-Public00613994446626
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,study
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date 12/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/14/2011
Device Model NumberD224TRK
Device Catalogue NumberD224TRK
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/19/2018
Initial Date FDA Received12/12/2018
Date Device Manufactured10/23/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0111-2011
Patient Sequence Number1
Treatment
418975C LEAD, 6945-65 LEAD, 6940-52 LEAD
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age62 YR
-
-