The icd was returned for analysis.The production process for this device was checked.The production documents did not show any irregularities.All production steps were correctly executed.A standard electrical outgoing goods test was performed and good working order was documented.Next, the device status was interrogated.Device status was bos and six charging procedures had been recorded.As described in the complaint, the icd was in safety (rom/backup) mode, as it was upon receipt.Generally speaking, the safety program is loaded from a read-only additional memory.As a result, the anti-bradycardia and anti-tachycardia parameters are pre-set in the safety program, and it is designed to ensure a level of safety that would cover the greatest possible portion of the patient population.The icd memory was examined.The investigation showed that the icd detected inconsistent memory content and, as a result, had automatically activated the safety (rom/backup) program on (b)(6) 2018.The device is generally able to self-correct individual memory sectors.If multiple memory sectors are affected at once, automatic correction is no longer possible and the device will respond as per specifications by switching to the safety mode to ensure patient safety.As per the clinical observation, the icd remained in the safety mode following re-initialization.To determine the root cause of this behavior, the icd was subjected to destructive analysis.The electronics module was sent to the manufacturer for an extensive components analysis.No visual irregularities were apparent.Following an extensive analysis of the individual components, the root cause for the clinical observation could be identified.Damage to the integrated circuit triggered the memory sectors in question.In summary, it can be assumed that the safety mode was activated due to the damaged integrated circuit.It can be assumed that the root cause of the clinical observation occurred after the device had been shipped.
|