HILL-ROM BATESVILLE VERSACARE FRAME; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P3200K000367
Device Problem Unintended System Motion (1430)
Patient Problem No Patient Involvement (2645)
Event Date 11/16/2018
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technical support representative performed troubleshooting over the phone with the account, asking the account to isolate the patient hand pendent.Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Head section motor: examine the motor for the presence and tightness of the attachment hardware.Replace as necessary.Fully raise and lower the head section.Make sure there is no friction or abnormal noises and no audible overload indication can be heard during the movement.Replace in the event of a malfunction.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.Hill-rom technical support contacted the account two additional times trying to obtain the resolution for this alleged issue.The account confirmed they repaired the bed, but they were unable to identify what action was taken to resolve the alleged issue.Based on this information, no further action is required.
 
Event Description
Hill-rom technical support received a report from the account stating the head section of the bed was raising on its own.The bed was located in 4 west at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
emily mitchell
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key8163139
MDR Text Key130481030
Report Number1824206-2018-00478
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberP3200K000367
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-