Catalog Number VS-402
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombus (2101)
Event Date 12/12/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Physician used a venaseal to treat the great saphenous vein (gsv).The ifu was followed.Local anaesthesia and hand compressions were performed.The physician was 5 cm away from the saphenofemoral junction junction (sfj) prior to treatment.The procedure was successfully completed resulting in vein closure.It was reported that a follow up ultrasound performed 48 hours after procedure showed a thrombus extension class 3 in the sfj.Patient had no symptoms.Anticoagulant therapy and zyralta was prescribed along follow up ultrasound in two weeks.
 
Manufacturer Narrative
Image review a single photograph of a sonographic image was provided.The sonographic image is from the area of the saphenofemoral junction and shows a thrombus extension of greater than 50% indicating that it is a level 3 extension.The event description mentions that the patient was placed on anticoagulant therapy and prescribe zyralta.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENA SEAL CLOSURE SYSTEM
Type of Device
AGENT, OCCLUDING, VASCULAR, PERMANENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key8163889
MDR Text Key130319643
Report Number9612164-2018-03626
Device Sequence Number1
Product Code PJQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberVS-402
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2018
Initial Date FDA Received12/14/2018
Supplement Dates Manufacturer Received03/05/2019
Supplement Dates FDA Received03/19/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient Weight59
-
-