Model Number 801763 |
Device Problem
Connection Problem (2900)
|
Patient Problems
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/04/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the distributor, every element was changed one by one, but the issue still occurred to anything connected to the right nic.The electronic patient gas system (epgs) was disconnected to check if that was the source of the problem, but the problem persisted.
|
|
Event Description
|
It was reported that while in the perfusion screen on the central control monitor (ccm), the elements connected to the right network interface card (nic) appear disconnected momentarily and an alarm sounds.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed via photos sent to the manufacturer.Multiple diligence attempts for part return were unsuccessful.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
Additional information received that the reported complaint occurred during preventive maintenance (pm).There was no patient involvement.
|
|