Model Number C3090907
Device Problems Material Separation (1562); Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/28/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation on-going.Supplemental report will be provided when additional information is available.
 
Event Description
It was reported that the suture thread spontaneously breaks.The frequency was unknown.The breakage occurs prior to use.Product was not used on a patient.
 
Manufacturer Narrative
Samples received: 4 unopened racepacks.Analysis and results: there are no previous complaints of this code-batch.We have received four closed samples to analyze this complaint.We have tested the knot pull tensile strength of the closed samples received and the results fulfil the requirements of the european pharmacopoeia (ep): 1.27 kgf in average and 1.17 kgf in minimum (ep requirements: 0.51 kgf in average and 0.15 kgf in minimum).Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfil usp/ep and b.Braun surgical requirements.Remarks: when working with optilene® suture material, great care should be taken to avoid any crushing or crimping damage of the monofilament by instruments such as forceps or needle holders.Final conclusion: although the results of the samples received fulfil the specifications of european pharmacopoeia/ b.Braun surgical specifications, we take note of this incidence in order to assess if new or additional actions are needed.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTILENE 4/0 (1,5) 90CM 2XHR26 CV RCP
Type of Device
SUTURE
Manufacturer (Section D)
B BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
MDR Report Key8164415
MDR Text Key130342756
Report Number3003639970-2018-00779
Device Sequence Number1
Product Code GAW
Combination Product (y/n)N
PMA/PMN Number
K133890
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 01/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/21/2023
Device Model NumberC3090907
Device Catalogue NumberC3090907
Device Lot Number118211
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2018
Distributor Facility Aware Date12/12/2018
Date Manufacturer Received12/12/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-