Physician was setting up a storz 8403sxi c-mac s video laryngoscope handle in preparation for use.While inserting the scope into the disposable laryngoscope blade he inadvertently inserted the handle approximately 180 degrees out of position causing the spring-loaded retainer on the end cap of the c-mac handle to become jammed in the top of the disposable blade.On attempting to remove the handle from the blade, the press-fit end cap of the c-mac handle became completely separated from the body of the handle, and the ribbon connector detached from the circuit board in the handle body.The c-mac handle was rendered unusable.Manufacturer response for video laryngoscope, c-mac s (per site reporter).Local clinical rep viewed the equipment in question, stated he was not aware of this failure happening elsewhere, and agreed that it should be replaced under warranty.
|