• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number MCP00703309
Device Problems Device Displays Incorrect Message (2591); Unintended Movement (3026)
Patient Problem No Patient Involvement (2645)
Event Date 12/03/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).
 
Event Description
As stated by the field service technician: ""i faced a problem which is related to rpm drive.After turning on the rpm drive, without any other action, the pump suddenly starts to rotate very fast and s-stop error is seen and the pump stops." no patient involvement.(b)(4).
 
Manufacturer Narrative
Maquet medical systems,usa(importer)submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary ag kehler strasse 31, 76437 rastatt, germany.A follow-up medwatch will be submitted when additional information becomes available.Reference exemption # e2018002.Importer- maquet medical systems usa (b)(4).Contact person- (b)(6).As stated by the field service technician the issue could be solved by replacing the motor control board.Despite several requests further information about service documents and part return have not been received.Thus no investigation could be performed.The root cause could not be determined.The failure could not be confirmed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.
 
Event Description
Internal reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HL-20 INTEGRATED PERFUSION SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
MDR Report Key8170314
MDR Text Key130786529
Report Number8010762-2018-00329
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 02/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMCP00703309
Device Catalogue Number70102.7652
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/03/2018
Initial Date FDA Received12/17/2018
Supplement Dates Manufacturer Received12/03/2018
Supplement Dates FDA Received02/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-