MAUDE Adverse Event Report: MPRI ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 3389-28
Device Problems Shipping Damage or Problem (1570); Material Deformation (2976); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/13/2018
Event Type  malfunction  
Manufacturer Narrative
Device was never implanted, and therefore never part of a system.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a user facility via a manufacturer representative (rep) regarding a patient who was implanted with a neurostimulator (ins) for unknown indications for use.The lead was twisted when removed from package by the physician.A new lead was used and the event resolved.There was no troubleshooting performed and no external factors noted.No patient harm was reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a health care provider (hcp) via a manufacturer representative (rep) regarding a patient who was implanted with a neurostimulator (ins) for unknown indications for use.The lead was twisted when removed from package by the physician.A new lead was used and the event resolved.There was no troubleshooting performed and no external factors noted.No patient harm was reported.On 12/21/2018, (for, hcp, rep): no new information.(corrected previous event description regarding initial reporter).
 
Manufacturer Narrative
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
The returned lead was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.The returned device passed all testing in the laboratory and no anomalies were identified.(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8172207
MDR Text Key130625448
Report Number2649622-2018-20792
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 05/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model Number3389-28
Device Catalogue Number3389-28
Device Lot NumberVA1TG22
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received05/07/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/31/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-