• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SCARBOROUGH INC. ALERE DETERMINE HIV 1/2 AG/AB COMBO; IVD FOR HIV AG/AB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SCARBOROUGH INC. ALERE DETERMINE HIV 1/2 AG/AB COMBO; IVD FOR HIV AG/AB Back to Search Results
Catalog Number 7D2648
Device Problem False Positive Result (1227)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
Testing was performed at alere (b)(4) on retained kit lot 102997 with the following internal serum plasma control samples: (b)(6).All test results were valid and performed as expected.Additionally, the manufacturing batch records for lot 102997 were reviewed.This lot met the required release specifications.A review of the complaints reported (b)(6) or unconfirmed (b)(6) related to lot number 102997 showed that the complaint rate is (b)(4).The evidence available does not indicate that the product is performing outside label claims.Alere (b)(4) was unable to determine the exact root cause of the reported issue.The results obtained may possibly be related to the patient sample.The sample may have contained specific substances which may have affected the results.The available evidence suggests that this device lot is performing within labeled claims.
 
Event Description
A customer reported (b)(6) results with the alere determine hiv 1/2 ag/ab combo on a serum sample.They further indicated that the same sample was tested with a different lot of alere determine hiv 1/2 ag/ab combo and produced (b)(6) results.The conflicting results suggest that a malfunction may have occurred.The patient gender, pregnancy status, treatment and patient outcome were unknown.The exact date of occurrence was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALERE DETERMINE HIV 1/2 AG/AB COMBO
Type of Device
IVD FOR HIV AG/AB
Manufacturer (Section D)
ALERE SCARBOROUGH INC.
10 southgate road
scarborough ME 04074
Manufacturer Contact
benjamin crystal
10 southgate road
scarborough, ME 04074
2077305820
MDR Report Key8173922
MDR Text Key131796510
Report Number1221359-2018-00556
Device Sequence Number1
Product Code MZF
UDI-Device Identifier10811877010293
UDI-Public011081187701029310102997
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BP120037-0
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/22/2020
Device Catalogue Number7D2648
Device Lot Number102997
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2018
Initial Date FDA Received12/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-