The hill-rom technician found the right side caregiver controls needed to be replaced.Per the hill-rom service manual, it is necessary for the centrella¿ bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Plug the bed into a power outlet.Make sure all functions on the caregiver control panel operate correctly.Repair or replace the siderail if necessary.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the right side caregiver controls to resolve the issue.Based on this information, no further action is required.
|