• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® ECLIPSE¿ SIGNAL¿ BLOOD COLLECTION NEEDLE WITH INTEGRATED HOLDER; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY (BD) BD VACUTAINER® ECLIPSE¿ SIGNAL¿ BLOOD COLLECTION NEEDLE WITH INTEGRATED HOLDER; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 368835
Device Problem Leak/Splash (1354)
Patient Problem Alteration In Body Temperature (2682)
Event Date 11/22/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for tube push off with the incident lot was not observed.Testing of the customer samples was performed and no tubes pushed off the needles.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Based on the investigation, a root cause could not be determined.The product was found to be in conformance and meet release specifications.
 
Event Description
It was reported with the use of the bd vacutainer® eclipse¿ signal¿ blood collection needle with integrated holder there was an issue with tube do not stay at holders sleeve, it flies away.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® ECLIPSE¿ SIGNAL¿ BLOOD COLLECTION NEEDLE WITH INTEGRATED HOLDER
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key8176473
MDR Text Key130892231
Report Number9617032-2018-03136
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeFI
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/31/2021
Device Catalogue Number368835
Device Lot Number8149794
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/29/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-