(b)(4).The implant or explant dates are not applicable to this device.Per the instructions for use (ifu) warns, never advance, torque or retract a pressure guide wire which meets significant resistance.The ifu also cautions, the pressure guide wire should not be advanced if resistance is encountered.The wire should never be forcibly pushed into a vessel.Any time that resistance is encountered, the wire should be withdrawn under fluoroscopic guidance.In some instances, the wire may kink and must be removed.The manufacturing documentation for this device was reviewed and the device met all quality and manufacturing release criteria.
|
This case was reviewed and investigated according to the manufacturer's policy.It was reported, during a diagnostic coronary procedure when inserting the manufacturer's device, it kinked with another manufacturer's catheter.The manufacturer's device was removed at the same time as another device.This event is being reported because the manufactures device and another manufacturer's guide catheter were removed as a system.
|