On (b)(6) 2018, after sizing the atrial septal defect using a 34mm sizing balloon and measuring the defect to 26mm, a 30mm amplatzer septal occluder was selected for implant.After deploying the device, a push-pull test was performed to confirm device stability and appeared to be well seated, observed on echocardiogram and angiogram.Then the device was released from the delivery cable and implanted.Following the procedure, while still in the hospital, the device was observed to have embolized to the right ventricle.The physician elected not to attempt snare or other percutaneous removal of the device as patient anatomy didn't allow for a larger device to close the defect.On (b)(6) 2018, the patient was transferred to another hospital where the device was surgically removed and the asd was patched successfully.Post-operatively, the patient is reported to be recovering.The user comments that the device embolized due to insufficient posterior rim tissue to hold the device in place and not related to the device.
|