• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. PULSE GEN MODEL 106; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problems Mechanical Problem (1384); Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/05/2018
Event Type  malfunction  
Event Description
It was reported that the generator that was intended to be implanted during the patient's replacement surgery presented a failure and therefore a different model 106 generator was implanted.The suspect product has not been received to date.No further relevant information has been received to date.
 
Event Description
It was reported that the device failure was that the setscrew was unable to be tightened by turning the hex screw driver.It was reported that the screw would not turn, met resistance.The device history records for the generator indicated that the generator passed all functional specifications prior to distribution.It was reported that the generator and the hex screwdriver were both discarded.No further relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
rachel kohn
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key8178620
MDR Text Key130898168
Report Number1644487-2018-02343
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public05425025750061
Combination Product (y/n)N
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/20/2020
Device Model Number106
Device Lot Number204518
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received01/11/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/15/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age13 YR
-
-