Model Number P1600B
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Patient Involvement (2645)
Event Date 12/07/2018
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technical support representative performed troubleshooting over the phone with the account, asking the account to isolate the side rails.Per the hill-rom service manual the advanta bed requires an effective maintenance program.Preventative maintenance will minimize downtime due to excessive wear.Run the bed to the full upper and lower limits to ensure proper function of both head and foot hilow drive limit switches.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The account replaced the right caregiver switch assembly to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the bed would run down after being plugged into ac power.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA BED GENERIC
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
emily mitchell
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key8186607
MDR Text Key131499395
Report Number1824206-2018-00489
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberP1600B
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/07/2018
Initial Date FDA Received12/20/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/29/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-