• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 TRI-LOCK BPS BROACH SZ 3; HIP INSTRUMENTS : BROACHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 TRI-LOCK BPS BROACH SZ 3; HIP INSTRUMENTS : BROACHES Back to Search Results
Catalog Number 201203030
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Code Available (3191)
Event Date 01/01/2018
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The size 3 trilock stems were not sitting at an acceptable depth relative to the size 3 trilock broach.After rebroaching the canal the stem still sat at a cm above the desired depth.
 
Manufacturer Narrative
Product complaint : (b)(4).Investigation summary : related dimensional inspection of the returned product found no discrepancy from drawing requirements.No product problem identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-LOCK BPS BROACH SZ 3
Type of Device
HIP INSTRUMENTS : BROACHES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 19380-0988
6103142063
MDR Report Key8186962
MDR Text Key131081801
Report Number1818910-2018-78925
Device Sequence Number1
Product Code HTQ
UDI-Device Identifier10603295082088
UDI-Public10603295082088
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number201203030
Device Lot NumberPG265583
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/27/2018
Initial Date FDA Received12/20/2018
Supplement Dates Manufacturer Received01/23/2019
Supplement Dates FDA Received01/23/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2017
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-