MAUDE Adverse Event Report: BAXTER HEALTHCARE ¿ ROUND LAKE 250 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE ¿ ROUND LAKE 250 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938737
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 11/26/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 250 ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag was leaking from the lower right corner on the seal of the bag.The leak was discovered prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was manufactured between 06/27/2018 - 06/28/2018.The actual device was received for evaluation.Visual inspection did not identify any issue with the seam of the bag in the lower right corner.However, visual inspection did observe a small tear in the lower left side edge of the bag.Functional testing was performed and a leak was observed from the tear in the lower left side of the bag only.Therefore, the reported issue of leak in the lower right corner was not verified, but a leak was verified in the lower left side of the bag.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
250 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE ¿ ROUND LAKE
round lake IL
Manufacturer (Section G)
BAXTER HEALTHCARE ¿ ROUND LAKE
25212 w. illinois route 120
round lake IL
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8187616
MDR Text Key131135006
Report Number1416980-2018-07958
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477213
UDI-Public(01)00085412477213
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 01/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2021
Device Catalogue NumberH938737
Device Lot Number60133469
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/28/2018
Initial Date FDA Received12/20/2018
Supplement Dates Manufacturer Received01/28/2019
Supplement Dates FDA Received01/30/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-