The device was manufactured between 06/27/2018 - 06/28/2018.The actual device was received for evaluation.Visual inspection did not identify any issue with the seam of the bag in the lower right corner.However, visual inspection did observe a small tear in the lower left side edge of the bag.Functional testing was performed and a leak was observed from the tear in the lower left side of the bag only.Therefore, the reported issue of leak in the lower right corner was not verified, but a leak was verified in the lower left side of the bag.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|