SMITH & NEPHEW, INC. HIP IMPLANT; PRSTHSIS, HIP, HMI-, TRNNION-BRING, FMRL, MTL/PLYCTL Back to Search Results
Catalog Number 71331750
Device Problems Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Swelling (2091); Injury (2348)
Event Date 12/13/2018
Event Type  Injury  
Event Description
It was reported that a revision surgery was performed because patient had pseudotumors.
 
Manufacturer Narrative
New information received upon investigation team approval.It was reported that a revision surgery was performed due to patient pseudotumors.The affected r3 3 hole acetabular shell, anthology porous stem, r3 32mm delta ceramic liner and ceramic head, used in treatment, were returned and evaluated.A lab analysis conducted during this investigation noted that on the femoral component gouge marks and an extensive biological on-growth in the proximal porous region are observed on the device.Marks most likely occurred during removal.This provides some indication of a well-fixed stem at the time of its removal.No physical damage or wear observed around the diameter of the locking taper.This implies the head was properly lock to the femoral component.The ceramic head did not reveal any physical damage but marks of metal transfer during removal are noted on the articular surface and lower side of head.The od porous surface of the shell exhibited biological on-growth, which suggests the shell was well-fixed at the time of revision.Dark scuff marks typical of metal transfer during implant revision are noted on the liner.From information presented and visual observations, there is no evidence to support any conclusion regarding the reported issue.A review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A relationship, if any, between the devices and the reported incident or adverse event could not be corroborated.No medical documents were received for investigation.Therefore no medical assessment can be performed at this time.Based on this investigation, the need for corrective action is not indicated.We will continue to monitor for future complaints and investigate as necessary.
 
Event Description
It was reported that a revision surgery was performed because patient had pseudotumors.During investigation of the device dark scuff marks were found on the liner.This was the only device were this wear was found.Pseudo tumour is offently linked to metal issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIP IMPLANT
Type of Device
PRSTHSIS, HIP, HMI-, TRNNION-BRING, FMRL, MTL/PLYCTL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
spa park
memphis TN 38116
MDR Report Key8190241
MDR Text Key131180116
Report Number3005975929-2018-00545
Device Sequence Number1
Product Code JDH
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number71331750
Device Lot Number11BT03637
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/13/2018
Initial Date FDA Received12/21/2018
Supplement Dates Manufacturer Received01/27/2020
Supplement Dates FDA Received02/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
P#71331950 L#11CM14870; P#71357004 L#10EM08802; P#UNKNOWN HIP IMPLANT L#UNKNOWN; UNKNOWN BHR DEVICE; UNKNOWN BHR DEVICE
Patient Outcome(s) Hospitalization; Required Intervention;
-
-