• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TNDM BP SHL/XLPE LNR 40OD 26ID; PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TNDM BP SHL/XLPE LNR 40OD 26ID; PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL Back to Search Results
Catalog Number 71325040
Device Problem Material Separation (1562)
Patient Problem Injury (2348)
Event Date 11/27/2018
Event Type  Injury  
Event Description
It was reported that revision surgery was performed.Locking ring was disassembled with the bipolar cup in vivo.Bipolar cup and head were revised.
 
Manufacturer Narrative
The associated devices were returned and evaluated.A dimensional analysis of the locking ring measured 4 dimensions are out of specification; od, id, groove diameter and spherical radius id.The bi-polar construct was returned unassembled.Minor scratches observed on the articulating surface of the bi-polar component.The femoral head reveal significant scratching/wear through the oxide layer on the articulating surface and a few scratches on the chamfer, near the female taper.It¿s unclear if the scratches occurred during removal, transport, in vivo or during implantation.The head is too large to pass through the opening of the bi-polar/liner construct when stacked.The head would need to pass through the opening in order to disassemble.This demonstrated that the head diameter was large enough to be constrained with the bi-polar component as designed.No deformation was visible on the locking ring, however the lateral surface was scratched significantly and a few minor scratches occurred.No deviations of material or manufacturing process are indicated for these components.No conclusions as to the cause of this issue can be determined.A clinical evaluation was conducted and based on the product evaluation during the extraction the implant, the head was too large to pull through the locking ring opening, oversizing numerous features and particularly the through-hole diameter.This is relatively common for this product.The deviations described in the product analysis are most likely due to extraction damage, and not the fault of the smith and nephew device itself.The future impact to the patient cannot be determined.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TNDM BP SHL/XLPE LNR 40OD 26ID
Type of Device
PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8192082
MDR Text Key131298212
Report Number1020279-2018-02877
Device Sequence Number1
Product Code JDH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number71325040
Device Lot Number15LM04417
Initial Date Manufacturer Received 11/28/2018
Initial Date FDA Received12/21/2018
Supplement Dates Manufacturer Received11/28/2018
Supplement Dates FDA Received05/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-