OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35; HF ELECTRODE, LOOP Back to Search Results
Model Number A22205C
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/05/2018
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to olympus for evaluation.The exact cause of the reported event could not be confirmed, however, based on similar reported complaints the reported event can occur from mechanical overload by the application of excessive force and or the electrode comes into contact (unintended) with other metal parts, e.G.Surgical instruments while the high-frequency output was activated.If the device is returned at a later date, this report will be supplemented accordingly.
 
Event Description
Olympus was informed that during a transurethral resection of a prostate (turp) procedure, the loop wire at the distal end of the device broke off and fell into the patient¿s bladder.The device was retrieved from the patient.There was no delay in the procedure as the intended procedure was completed using a second device from the same lot.In addition, there was no sparking or arcing observed during procedure; there was no unexpected bleeding to the patient; there was no issue withdrawing the device from the patient and the patient did not require additional procedures or treatment.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the device evaluation results.A visual inspection was performed on the received condition and confirmed the loop wire at the distal end of the device was broken and missing.A microscope inspection revealed char marks and melting at the breakage point on both yellow colored insulation posts at the distal tip.The two insulation posts at the distal section were bent inward; the red insulation shaft is straight, and its proximal end has no physical damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35
Type of Device
HF ELECTRODE, LOOP
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
MDR Report Key8192363
MDR Text Key131772804
Report Number2951238-2018-00801
Device Sequence Number1
Product Code GCP
UDI-Device Identifier14042761036709
UDI-Public14042761036709
Combination Product (y/n)N
PMA/PMN Number
K790071
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 02/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberA22205C
Device Lot Number1000024185
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-