Model Number CD3357-40C |
Device Problem
Over-Sensing (1438)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/19/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that when patient presented remotely via merlin.Net transmission, multiple auto mode switch episodes were observed over multiple days.This appeared to be due to far field r-wav oversensing.Programming changes were recommended which will be performed during the next scheduled follow up visit.Patient was stable.
|
|
Event Description
|
New information received on december 21st, 2018 states that the recommended programming changes were not made yet and will be made during the patient¿s next in clinic visit.
|
|
Event Description
|
New information received on february 8th, 2019 states that the recommended programming changes were made.
|
|