• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA (DDQP+); IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA (DDQP+); IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3265-40Q
Device Problem Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/16/2018
Event Type  Injury  
Manufacturer Narrative
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
 
Event Description
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and awaiting explant.Further information was requested, but not yet available.
 
Event Description
New information on (b)(6) 2019 stated that on (b)(6) 2019, the device was explanted and replaced.Patient was fine.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.A device evaluation was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA (DDQP+)
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8192495
MDR Text Key131299658
Report Number2017865-2018-19453
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734505673
UDI-Public05414734505673
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 04/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2014
Device Model NumberCD3265-40Q
Device Catalogue NumberCD3265-40Q
Device Lot Number3818056
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/24/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/27/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0004-2018; Z-0117-2017
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-