Catalog Number RTLR180343
Device Problem Overfill (2404)
Patient Problems Memory Loss/Impairment (1958); Uremia (2188); Weight Changes (2607)
Event Date 11/12/2018
Event Type  Injury  
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.Clinical investigation: there is a temporal relationship between pd therapy on the liberty select cycler and the patient event of fluid volume overload with uremia and cognitive impairment.The uremia which is due to insufficient dialysis is most likely the cause of the patient¿s confusion and memory issues.There is a possible causal relationship between the liberty select cycler and the adverse event as the patient reported unspecified issues with the liberty select cycler which prevented him from completing ccpd therapy.A machine evaluation has not been completed at the time of this clinical investigation and there are no treatment records for review, therefore it cannot be confirmed if any malfunction occurred.Based on the available information the cause of the event cannot be confirmed.
 
Event Description
On (b)(6) 2018 during follow up for a separate issue, the peritoneal dialysis registered nurse (pdrn) for this male patient on continuous cycling peritoneal dialysis (ccpd) for renal replacement therapy (rrt) reported the patient was hospitalized (b)(6) 2018.The patient was hospitalized for fluid volume overload with a weight gain of seventeen pounds (unknown time period) with uremia, concentration issues and short-term memory loss.The patient had stated they were having unspecified issues with the fresenius liberty select cycler and could not get past dwell three.The patient was discharged (b)(6) 2018.The pdrn reported that the cycler issues could not be confirmed.The hospital course is unknown.
 
Manufacturer Narrative
Changed product problem to yes.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key8195472
MDR Text Key131373349
Report Number2937457-2018-03725
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861102068
UDI-Public00840861102068
Combination Product (y/n)N
PMA/PMN Number
K171652
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 12/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberRTLR180343
Was Device Available for Evaluation? No
Device Age MO
Initial Date Manufacturer Received 12/26/2018
Initial Date FDA Received12/26/2018
Supplement Dates Manufacturer Received12/26/2018
Supplement Dates FDA Received12/26/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DELFLEX PD FLUID; LIBERTY CYCLER SET; DELFLEX PD FLUID; LIBERTY CYCLER SET
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age53 YR
Patient Weight83
-
-