• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-032
Device Problem Unintended Movement (3026)
Patient Problem No Information (3190)
Event Date 11/27/2018
Event Type  Injury  
Manufacturer Narrative
An event of embolization was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined; however, information from the field indicated that the user believed this was most likely due to mis-sizing.
 
Event Description
On (b)(6) 2018, a 30mm amplatzer atrial septal occluder (aso) was selected.During the procedure, the 30mm aso deployed in a cobra shape and per the user, the deformation event did not have the potential to cause patient harm or injury.The device was exchanged for a second 30mm aso, which was considered a mis-sizing so the device was exchanged for a 32mm aso.The 32mm aso was deployed and released, but embolized after implantation.Again, per user, the embolization event was most likely due to mis-sizing of the device.The patient was referred to cardiac surgery where the embolized device was retrieved and the atrial septal defect was closed with a patch.The patient remained hemodynamically stable throughout the procedure and is reported stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8196075
MDR Text Key131394674
Report Number2135147-2018-00246
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010236
UDI-Public00811806010236
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model Number9-ASD-032
Device Catalogue Number9-ASD-032
Device Lot Number6603642
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/28/2018
Initial Date FDA Received12/26/2018
Date Device Manufactured08/23/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age22 YR
-
-