The hill-rom technician found the external alarm needed to be reconnected.Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Make sure the brake is not set, and then plug the bed into an applicable power source.A steady alarm comes on.Set the brake.The alarm stops.Repair or replace as necessary.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in november 2018.It is unknown if the facility performed any other preventative maintenance on this bed.The technician reconnected the external alarm to resolve the issue.Based on this information, no further action is required.
|