• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-PFO-025
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/04/2018
Event Type  Injury  
Event Description
On (b)(6) 2018, a 25mm amplatzer pfo was selected for use.During deployment, the device would not sit flat on the septum and was removed.The 25mm pfo was observed bulbous outside the body and then looked normal shortly after.A second 25mm amplatzer pfo was implanted without issue.There were no known complications with this procedure and the patient remained hemodynamically stable with no harm or injury.
 
Manufacturer Narrative
The reported event of deformity upon deployment could not be confirmed.The investigation confirmed the device met functional specifications when analyzed at abbott under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER PFO OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8196743
MDR Text Key131415104
Report Number2135147-2018-00253
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806011288
UDI-Public00811806011288
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model Number9-PFO-025
Device Catalogue Number9-PFO-025
Device Lot Number6707461
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/04/2018
Initial Date FDA Received12/26/2018
Supplement Dates Manufacturer Received01/23/2019
Supplement Dates FDA Received01/24/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight100
-
-