• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 10623
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/14/2018
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.Device is a combination product.
 
Event Description
It was reported that a stent damage occurred.The 75% stenosed target lesion was located in the mildly tortuous and severely calcified left anterior descending artery.An imaging catheter for intravascular ultrasound and a 2.50 x 38mm synergy drug-eluting stent were advanced but failed to cross the lesion.When the stent was removed from the patient's body, the stent struts were found to be lifted.The procedure was completed using a non- bsc device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8197845
MDR Text Key131476988
Report Number2134265-2018-64649
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/07/2020
Device Model Number10623
Device Catalogue Number10623
Device Lot Number0022390989
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2018
Initial Date FDA Received12/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/11/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-