• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problems Break (1069); Loose or Intermittent Connection (1371); Failure to Power Up (1476); Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/07/2018
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of battery charger/modem sn (b)(4) has been completed.The reported problem (unable to power on, damaged power supply connector) has been confirmed.Upon investigation the battery charger/modem was unable to power on and recharge a battery.The cause for the failure was isolated to a damaged power supply connector.The root cause for the damaged power supply connector was excessive force.No adverse event resulted from the defective battery charger/modem.
 
Event Description
A us distributor reported that a patient's battery charger was unable to power on and that the power supply connector was damaged/loose.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
brooke arnold
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key8200356
MDR Text Key131712842
Report Number3008642652-2018-10811
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005029
UDI-Public00855778005029
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 12/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2018
Initial Date Manufacturer Received 12/26/2018
Initial Date FDA Received12/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-