ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, DUAL 4 CHANNEL 10CM; SCS EXTENSION Back to Search Results
Model Number 3341
Device Problems Disconnection (1171); High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Date 12/03/2018
Event Type  Injury  
Event Description
Device 2 of 3: reference mfr report#3006705815-2018-03459, reference mfr report#1627487-2018-13405.It was reported that the lead diagnostics revealed high impedances on the contacts.X rays confirmed extension was pulled from the header of the ipg.As a result, surgical intervention was undertaken on (b)(6) 2018, wherein the extensions were replaced electively because the lead was disconnected and fluid invasion was a concern.Reportedly, stimulation was restored post-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, DUAL 4 CHANNEL 10CM
Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8201013
MDR Text Key131580058
Report Number1627487-2018-13404
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05414734402279
UDI-Public05414734402279
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K930536
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/25/2020
Device Model Number3341
Device Lot Number6378772
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2018
Initial Date FDA Received12/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/26/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age31 YR
-
-